Conventional linear accelerators (LINACs) for radiotherapy produce fast secondary neutrons due to photonuclear processes. The neutron presence is considered as an extra undesired dose during the radiotherapy treatment, which could cause secondary radio-induced tumors and malfunctions to cardiological implantable devices. It is thus important to measure the neutron dose contribution to patients during radiotherapy, not only at high-energy LINACs, but also at lower energies, near the giant dipole resonance reaction threshold. In this work, the full body neutron dose equivalent has been measured during single-field radiotherapy sessions carried out at different LINAC energies (15, 10 and 6 MV) by using a tissue equivalent (for neutrons) anthropomorphic phantom together with bubble dosemeters. Results have shown that some neutron photoproduction is still present also at lower energies. As a consequence, emitted photoneutrons cannot be ignored and represent a risk contribution for patients undergoing radiotherapy.

MEASUREMENTS OF THE PARASITIC NEUTRON DOSE AT ORGANS FROM MEDICAL LINACS AT DIFFERENT ENERGIES BY USING BUBBLE DETECTORS

Katia Alikaniotis
;
Mara Severgnini;Gianrossano Giannini;Vittorino Milan
2018-01-01

Abstract

Conventional linear accelerators (LINACs) for radiotherapy produce fast secondary neutrons due to photonuclear processes. The neutron presence is considered as an extra undesired dose during the radiotherapy treatment, which could cause secondary radio-induced tumors and malfunctions to cardiological implantable devices. It is thus important to measure the neutron dose contribution to patients during radiotherapy, not only at high-energy LINACs, but also at lower energies, near the giant dipole resonance reaction threshold. In this work, the full body neutron dose equivalent has been measured during single-field radiotherapy sessions carried out at different LINAC energies (15, 10 and 6 MV) by using a tissue equivalent (for neutrons) anthropomorphic phantom together with bubble dosemeters. Results have shown that some neutron photoproduction is still present also at lower energies. As a consequence, emitted photoneutrons cannot be ignored and represent a risk contribution for patients undergoing radiotherapy.
2018
1-ago-2018
Pubblicato
https://academic.oup.com/rpd/article/180/1-4/267/4816199
File in questo prodotto:
File Dimensione Formato  
ncx308.pdf

Accesso chiuso

Tipologia: Documento in Versione Editoriale
Licenza: Copyright Editore
Dimensione 305.59 kB
Formato Adobe PDF
305.59 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11368/2928336
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 1
social impact