RICH-1 is a Cherenkov Imaging detector in operation at the COMPASS experiment at CERN SPS since 2001. MWPCs equipped with CsI photocathodes have been used as photon detectors, for the full detection area in the first years of operation, and for 75% of the area after a detector upgrade in 2006. The mean number of detected photons per particle has been extracted from the data collected during six years of COMPASS data taking. An indirect measurement of possible variations of the CsI quantum efficiency versus time is obtained from this analysis.
Long term experience with CsI photocathodes in gas photon detectors
BIRSA, RENATO;BRADAMANTE, FRANCO;BRESSAN, Andrea;DUIC, VENICIO;GIORGI, MARCELLO;LEVORATO, STEFANO;MARTIN, ANNA;SBRIZZAI, GIULIO;SCHIAVON, PAOLO;SOZZI, FEDERICA;
2014-01-01
Abstract
RICH-1 is a Cherenkov Imaging detector in operation at the COMPASS experiment at CERN SPS since 2001. MWPCs equipped with CsI photocathodes have been used as photon detectors, for the full detection area in the first years of operation, and for 75% of the area after a detector upgrade in 2006. The mean number of detected photons per particle has been extracted from the data collected during six years of COMPASS data taking. An indirect measurement of possible variations of the CsI quantum efficiency versus time is obtained from this analysis.File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.