Purpose: To compare the intraoperative parameters and postoperative outcomes using two different phaco handpieces with different energy delivery systems of the same phacoemulsification machine. Methods: One hundred and sixty eyes of 160 cataract patients were randomly assigned to 2.40 mm phacoemulsification using the AMO WhiteStar Signature® phaco system with the standard phaco handpiece (Group 1) or the Ellips FX phaco handpiece (Group 2). According to the lens opacities classification system III, each group was divided into four subgroups of 20 patients each. U/S total time (UST), total phaco time (TPT), mean effective phaco time, cumulative dissipated energy (CDE) and balanced salt solution (BSS) consumption were studied. Visual acuity, corneal endothelium cell count and corneal pachymetry were evaluated before surgery and 1, 7 and 30 days after surgery. Results: No statistically significant difference in terms of visual acuity was observed between the two groups. The postoperative endothelial cell loss and corneal pachymetry were similar between the two groups. In higher density cataracts, Ellips FX showed the lower CDE and the lowest UST. TPT and BSS consumption were significantly lower in the Ellips FX handpiece group. Conclusions: The two phaco handpieces appeared to be safe. AMO Ellips FX handpiece showed better performances and may be safer in lens removal, especially in hard cataract patients.

Comparative clinical study of Whitestar Signature phacoemulsification system with standard and Ellips FX handpieces

TOGNETTO, DANIELE
;
D'ALOISIO, ROSSELLA;CECCHINI, PAOLO;DI MARTINO, GIUSEPPE
2018-01-01

Abstract

Purpose: To compare the intraoperative parameters and postoperative outcomes using two different phaco handpieces with different energy delivery systems of the same phacoemulsification machine. Methods: One hundred and sixty eyes of 160 cataract patients were randomly assigned to 2.40 mm phacoemulsification using the AMO WhiteStar Signature® phaco system with the standard phaco handpiece (Group 1) or the Ellips FX phaco handpiece (Group 2). According to the lens opacities classification system III, each group was divided into four subgroups of 20 patients each. U/S total time (UST), total phaco time (TPT), mean effective phaco time, cumulative dissipated energy (CDE) and balanced salt solution (BSS) consumption were studied. Visual acuity, corneal endothelium cell count and corneal pachymetry were evaluated before surgery and 1, 7 and 30 days after surgery. Results: No statistically significant difference in terms of visual acuity was observed between the two groups. The postoperative endothelial cell loss and corneal pachymetry were similar between the two groups. In higher density cataracts, Ellips FX showed the lower CDE and the lowest UST. TPT and BSS consumption were significantly lower in the Ellips FX handpiece group. Conclusions: The two phaco handpieces appeared to be safe. AMO Ellips FX handpiece showed better performances and may be safer in lens removal, especially in hard cataract patients.
2018
13-lug-2017
Pubblicato
File in questo prodotto:
File Dimensione Formato  
Tognetto2018_Article_ComparativeClinicalStudyOfWhit.pdf

Accesso chiuso

Tipologia: Documento in Versione Editoriale
Licenza: Copyright Editore
Dimensione 594.18 kB
Formato Adobe PDF
594.18 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Tognetto2018_Article_ComparativeClinicalStudyOfWhit-Post_print.pdf

Open Access dal 14/07/2018

Tipologia: Bozza finale post-referaggio (post-print)
Licenza: Digital Rights Management non definito
Dimensione 936.79 kB
Formato Adobe PDF
936.79 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11368/2907218
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 4
social impact