Breast computed tomography (BCT) is an emerging application of X-ray tomography in radiological practice. A few clinical prototypes are under evaluation in hospitals and new systems are under development aiming at improving spatial and contrast resolution and reducing delivered dose. At the same time, synchrotron-radiation phase-contrast mammography has been demonstrated to offer substantial advantages when compared with conventional mammography. At Elettra, the Italian synchrotron radiation facility, a clinical program of phase-contrast BCT based on the free-space propagation approach is under development. In this paper, full-volume breast samples imaged with a beam energy of 32 keV delivering a mean glandular dose of 5 mGy are presented. The whole acquisition setup mimics a clinical study in order to evaluate its feasibility in terms of acquisition time and image quality. Acquisitions are performed using a high-resolution CdTe photon-counting detector and the projection data are processed via a phase-retrieval algorithm. Tomographic reconstructions are compared with conventional mammographic images acquired prior to surgery and with histologic examinations. Results indicate that BCT with monochromatic beam and free-space propagation phase-contrast imaging provide relevant three-dimensional insights of breast morphology at clinically acceptable doses and scan times.
Advancements towards the implementation of clinical phase-contrast breast computed tomography at Elettra / Longo, Renata; Arfelli, Fulvia; Bonazza, Deborah; Bottigli, Ubaldo; Brombal, Luca; Contillo, Adriano; Cova, Maria A.; Delogu, Pasquale; Di Lillo, Francesca; Di Trapani, Vittorio; Donato, Sandro; Dreossi, Diego; Fanti, Viviana; Fedon, Christian; Golosio, Bruno; Mettivier, Giovanni; Oliva, Piernicola; Pacilè, Serena; Sarno, Antonio; Rigon, Luigi; Russo, Paolo; Taibi, Angelo; Tonutti, Maura; Zanconati, Fabrizio; Tromba, Giuliana. - In: JOURNAL OF SYNCHROTRON RADIATION. - ISSN 1600-5775. - 26:4(2019), pp. 1343-1353. [10.1107/S1600577519005502]
Advancements towards the implementation of clinical phase-contrast breast computed tomography at Elettra
Longo, Renata;Arfelli, Fulvia;Bonazza, Deborah;Bottigli, Ubaldo;Brombal, Luca
;Cova, Maria A.;Di Trapani, Vittorio;Donato, Sandro;Dreossi, Diego;Fedon, Christian;Pacilè, Serena;Rigon, Luigi;Taibi, Angelo;Zanconati, Fabrizio;
2019-01-01
Abstract
Breast computed tomography (BCT) is an emerging application of X-ray tomography in radiological practice. A few clinical prototypes are under evaluation in hospitals and new systems are under development aiming at improving spatial and contrast resolution and reducing delivered dose. At the same time, synchrotron-radiation phase-contrast mammography has been demonstrated to offer substantial advantages when compared with conventional mammography. At Elettra, the Italian synchrotron radiation facility, a clinical program of phase-contrast BCT based on the free-space propagation approach is under development. In this paper, full-volume breast samples imaged with a beam energy of 32 keV delivering a mean glandular dose of 5 mGy are presented. The whole acquisition setup mimics a clinical study in order to evaluate its feasibility in terms of acquisition time and image quality. Acquisitions are performed using a high-resolution CdTe photon-counting detector and the projection data are processed via a phase-retrieval algorithm. Tomographic reconstructions are compared with conventional mammographic images acquired prior to surgery and with histologic examinations. Results indicate that BCT with monochromatic beam and free-space propagation phase-contrast imaging provide relevant three-dimensional insights of breast morphology at clinically acceptable doses and scan times.| File | Dimensione | Formato | |
|---|---|---|---|
|
Longo_et_al-2019-Journal_of_Synchrotron_Radiation.pdf
Accesso chiuso
Descrizione: Articolo principale
Tipologia:
Documento in Versione Editoriale
Licenza:
Copyright Editore
Dimensione
3.39 MB
Formato
Adobe PDF
|
3.39 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
|
2944840_Longo_et_al-2019-Journal_of_Synchrotron_Radiation-PostPrint.pdf
accesso aperto
Descrizione: PostPrint VQR3
Tipologia:
Bozza finale post-referaggio (post-print)
Licenza:
Digital Rights Management non definito
Dimensione
3.32 MB
Formato
Adobe PDF
|
3.32 MB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.


