The aim of this study is to describe an alternative technique to secure the receiver-stimulator of the cochlear implant to the skull with a tailored flap of periosteum. Other techniques are also reviewed and discussed. 179 consecutive patients were implanted by the same surgeon in a tertiary care setting. Age ranged from 11 months to 74 years. Patients were retrospectively evaluated for device migration. No cases of migration were observed during follow up, ranged 1-99 months with a median of 48 months. The alternative technique proposed is safe and reliable.
Non-sutured fixation of the internal receiver-stimulator in cochlear implantation / BOSCOLO-RIZZO, P., Muzzi, E., Trabalzini, F.. - In: EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY. - ISSN 0937-4477. - STAMPA. - 268:(2011), pp. 961-965. [10.1007/s00405-010-1479-5]
Non-sutured fixation of the internal receiver-stimulator in cochlear implantation
P. BOSCOLO-RIZZO;
2011-01-01
Abstract
The aim of this study is to describe an alternative technique to secure the receiver-stimulator of the cochlear implant to the skull with a tailored flap of periosteum. Other techniques are also reviewed and discussed. 179 consecutive patients were implanted by the same surgeon in a tertiary care setting. Age ranged from 11 months to 74 years. Patients were retrospectively evaluated for device migration. No cases of migration were observed during follow up, ranged 1-99 months with a median of 48 months. The alternative technique proposed is safe and reliable.Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.