Purpose: to investigate the reproducibility and reliability of OCT-A vascular measurements using Heidelberg Spectralis II OCT-A. Methods: a prospective study involving a single eye of patients aged 18 or older with no ocular disease. In order to investigate the reliability of the first and second OCT-A scans, the coefficient of variation of the foveal avascular zone (FAZ) and the vessel density (VD) in the superficial (SCP), intermediate (ICP) and deep capillary plexus (DCP) were calculated. Results: A total of 75 eyes were included in the study. The mean FAZ in the first and second scan was 0.36 × 0.13 mm2 and 0.37 × 0.12 mm2, respectively, in the SCP, 0.23 × 0.10 mm2 and 0.23 × 0.09 mm2 in the ICP, and 0.42 × 0.11 mm2 and 0.43 × 0.12 mm2 in the DCP. The overall VD was 36.05 × 9.01 and 35.33 × 9.92 at the first and second scan, respectively, in the SCP, 21.87 × 5.00 and 21.32 × 5.56 in the ICP, and 23.84 × 6.53 and 23.20 × 6.83 in the DCP. No statistically significant differences in FAZ measurements and VD in all sectors of each capillary plexus were observed between the first and second scan (p > 0.05). Conclusion: our study demonstrated the good reproducibility and reliability of OCT-A vascular measurements in the analysis of the FAZ and the quantification of VD in each capillary plexus of the retina.
Reproducibility and Reliability of Spectralis II OCT Angiography Vascular Measurements
Pastore, Marco R.
;Grotto, Alberto;Vezzoni, Francesca;Gaggino, Andrea;Milan, Serena;Gouigoux, Stefano;Guerin, Pier Luigi;Cirigliano, Gabriella;Tognetto, Daniele
2022-01-01
Abstract
Purpose: to investigate the reproducibility and reliability of OCT-A vascular measurements using Heidelberg Spectralis II OCT-A. Methods: a prospective study involving a single eye of patients aged 18 or older with no ocular disease. In order to investigate the reliability of the first and second OCT-A scans, the coefficient of variation of the foveal avascular zone (FAZ) and the vessel density (VD) in the superficial (SCP), intermediate (ICP) and deep capillary plexus (DCP) were calculated. Results: A total of 75 eyes were included in the study. The mean FAZ in the first and second scan was 0.36 × 0.13 mm2 and 0.37 × 0.12 mm2, respectively, in the SCP, 0.23 × 0.10 mm2 and 0.23 × 0.09 mm2 in the ICP, and 0.42 × 0.11 mm2 and 0.43 × 0.12 mm2 in the DCP. The overall VD was 36.05 × 9.01 and 35.33 × 9.92 at the first and second scan, respectively, in the SCP, 21.87 × 5.00 and 21.32 × 5.56 in the ICP, and 23.84 × 6.53 and 23.20 × 6.83 in the DCP. No statistically significant differences in FAZ measurements and VD in all sectors of each capillary plexus were observed between the first and second scan (p > 0.05). Conclusion: our study demonstrated the good reproducibility and reliability of OCT-A vascular measurements in the analysis of the FAZ and the quantification of VD in each capillary plexus of the retina.File | Dimensione | Formato | |
---|---|---|---|
Reproducibility and Reliability of Spectralis II OCT.pdf
accesso aperto
Tipologia:
Documento in Versione Editoriale
Licenza:
Creative commons
Dimensione
2.84 MB
Formato
Adobe PDF
|
2.84 MB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.