Purpose: to investigate the reproducibility and reliability of OCT-A vascular measurements using Heidelberg Spectralis II OCT-A. Methods: a prospective study involving a single eye of patients aged 18 or older with no ocular disease. In order to investigate the reliability of the first and second OCT-A scans, the coefficient of variation of the foveal avascular zone (FAZ) and the vessel density (VD) in the superficial (SCP), intermediate (ICP) and deep capillary plexus (DCP) were calculated. Results: A total of 75 eyes were included in the study. The mean FAZ in the first and second scan was 0.36 × 0.13 mm2 and 0.37 × 0.12 mm2, respectively, in the SCP, 0.23 × 0.10 mm2 and 0.23 × 0.09 mm2 in the ICP, and 0.42 × 0.11 mm2 and 0.43 × 0.12 mm2 in the DCP. The overall VD was 36.05 × 9.01 and 35.33 × 9.92 at the first and second scan, respectively, in the SCP, 21.87 × 5.00 and 21.32 × 5.56 in the ICP, and 23.84 × 6.53 and 23.20 × 6.83 in the DCP. No statistically significant differences in FAZ measurements and VD in all sectors of each capillary plexus were observed between the first and second scan (p > 0.05). Conclusion: our study demonstrated the good reproducibility and reliability of OCT-A vascular measurements in the analysis of the FAZ and the quantification of VD in each capillary plexus of the retina.
Reproducibility and Reliability of Spectralis II OCT Angiography Vascular Measurements / Pastore, Marco R.; Grotto, Alberto; Vezzoni, Francesca; Gaggino, Andrea; Milan, Serena; Gouigoux, Stefano; Guerin, Pier Luigi; Vinguerra, Alex L.; Cirigliano, Gabriella; Tognetto, Daniele. - In: DIAGNOSTICS. - ISSN 2075-4418. - ELETTRONICO. - 12:8(2022), pp. 1908."-"-1908."-". [10.3390/diagnostics12081908]
Reproducibility and Reliability of Spectralis II OCT Angiography Vascular Measurements
Pastore, Marco R.
;Grotto, Alberto;Vezzoni, Francesca;Gaggino, Andrea;Milan, Serena;Gouigoux, Stefano;Guerin, Pier Luigi;Cirigliano, Gabriella;Tognetto, Daniele
2022-01-01
Abstract
Purpose: to investigate the reproducibility and reliability of OCT-A vascular measurements using Heidelberg Spectralis II OCT-A. Methods: a prospective study involving a single eye of patients aged 18 or older with no ocular disease. In order to investigate the reliability of the first and second OCT-A scans, the coefficient of variation of the foveal avascular zone (FAZ) and the vessel density (VD) in the superficial (SCP), intermediate (ICP) and deep capillary plexus (DCP) were calculated. Results: A total of 75 eyes were included in the study. The mean FAZ in the first and second scan was 0.36 × 0.13 mm2 and 0.37 × 0.12 mm2, respectively, in the SCP, 0.23 × 0.10 mm2 and 0.23 × 0.09 mm2 in the ICP, and 0.42 × 0.11 mm2 and 0.43 × 0.12 mm2 in the DCP. The overall VD was 36.05 × 9.01 and 35.33 × 9.92 at the first and second scan, respectively, in the SCP, 21.87 × 5.00 and 21.32 × 5.56 in the ICP, and 23.84 × 6.53 and 23.20 × 6.83 in the DCP. No statistically significant differences in FAZ measurements and VD in all sectors of each capillary plexus were observed between the first and second scan (p > 0.05). Conclusion: our study demonstrated the good reproducibility and reliability of OCT-A vascular measurements in the analysis of the FAZ and the quantification of VD in each capillary plexus of the retina.| File | Dimensione | Formato | |
|---|---|---|---|
|
Reproducibility and Reliability of Spectralis II OCT.pdf
accesso aperto
Tipologia:
Documento in Versione Editoriale
Licenza:
Creative commons
Dimensione
2.84 MB
Formato
Adobe PDF
|
2.84 MB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.