The Timepix4 is an application-specific integrated circuit developed by the Medipix4 collaboration. The chip is a 448 x 512 matrix that can be connected, via bump-bonding, to pixelated sensors of various materials and thicknesses, forming application specific detection systems. Among the Timepix4 operating modes, the data-driven Time-of-Arrival - Time-Over-Threshold is a particularly promising mode for spectral imaging applications: data packets are generated every time the charge created by the incoming radiation goes above a pre-defined threshold, allowing the count of the detected particles and their energy measurement. This article investigates the spectroscopic characteristics of a Timepix4 assembly equipped with a 300 mu m Si sensor, using monochromatic X-ray beams in the energy range 8.5 keV-40 keV working at the SYRMEP beamline of the Elettra synchrotron. The accuracy and the energy resolution of the reconstructed energy spectra, together with an energy resolution model, will be presented.
Validation of Timepix4 energy calibration procedures with synchrotron X-ray beams
Brombal, L.;Brun, F.;Cardarelli, P.;Longo, R.;
2024-01-01
Abstract
The Timepix4 is an application-specific integrated circuit developed by the Medipix4 collaboration. The chip is a 448 x 512 matrix that can be connected, via bump-bonding, to pixelated sensors of various materials and thicknesses, forming application specific detection systems. Among the Timepix4 operating modes, the data-driven Time-of-Arrival - Time-Over-Threshold is a particularly promising mode for spectral imaging applications: data packets are generated every time the charge created by the incoming radiation goes above a pre-defined threshold, allowing the count of the detected particles and their energy measurement. This article investigates the spectroscopic characteristics of a Timepix4 assembly equipped with a 300 mu m Si sensor, using monochromatic X-ray beams in the energy range 8.5 keV-40 keV working at the SYRMEP beamline of the Elettra synchrotron. The accuracy and the energy resolution of the reconstructed energy spectra, together with an energy resolution model, will be presented.File | Dimensione | Formato | |
---|---|---|---|
Delogu et al. - 2024 - Validation of Timepix4 energy calibration procedures with synchrotron X-ray beams.pdf
accesso aperto
Tipologia:
Documento in Versione Editoriale
Licenza:
Creative commons
Dimensione
2.04 MB
Formato
Adobe PDF
|
2.04 MB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.