The anticancer drug imatinib is often involved in therapeutic drug monitoring (TDM) studies aimed at improving the treatment of several forms of leukemia and gastrointestinal stromal tumors (GIST). To further implement the TDM of imatinib in clinical practice, we developed a detection assay by using an ssDNA aptamer, which demonstrated excellent selectivity and was not affected by interference from the components of human plasma samples. The efficient binding of imatinib to the aptamer was demonstrated by means of surface plasmon resonance (SPR) analysis, which allowed the development of a quantitative assay in the concentration range between 400 and 6000 ng mL-1 (0.7-10 μM), where a lower limit of quantification (LLOQ) of 400 ng mL-1 was achieved. The precision of the assay was found to be within 12.0%, whereas the accuracy was in a range between 97.1 and 101.5%. The sample preparation procedure displayed a recovery in the range of 48.8-52.8%. Solid validation data were collected according to the regulatory guidelines and the method was compared with standard analytical techniques, leading to the development of a feasible aptasensor for the TDM of patients administered with imatinib.

An SPR investigation into the therapeutic drug monitoring of the anticancer drug imatinib with selective aptamers operating in human plasma

Bellotto O.;Poetto A. S.;Zanchetta M.;Posocco B.;Toffoli G.
2020-01-01

Abstract

The anticancer drug imatinib is often involved in therapeutic drug monitoring (TDM) studies aimed at improving the treatment of several forms of leukemia and gastrointestinal stromal tumors (GIST). To further implement the TDM of imatinib in clinical practice, we developed a detection assay by using an ssDNA aptamer, which demonstrated excellent selectivity and was not affected by interference from the components of human plasma samples. The efficient binding of imatinib to the aptamer was demonstrated by means of surface plasmon resonance (SPR) analysis, which allowed the development of a quantitative assay in the concentration range between 400 and 6000 ng mL-1 (0.7-10 μM), where a lower limit of quantification (LLOQ) of 400 ng mL-1 was achieved. The precision of the assay was found to be within 12.0%, whereas the accuracy was in a range between 97.1 and 101.5%. The sample preparation procedure displayed a recovery in the range of 48.8-52.8%. Solid validation data were collected according to the regulatory guidelines and the method was compared with standard analytical techniques, leading to the development of a feasible aptasensor for the TDM of patients administered with imatinib.
2020
11-dic-2020
Pubblicato
File in questo prodotto:
File Dimensione Formato  
Analyst_Cro Aviano_2020_complete.pdf

Accesso chiuso

Tipologia: Documento in Versione Editoriale
Licenza: Copyright Editore
Dimensione 1.42 MB
Formato Adobe PDF
1.42 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
d0an01860k1.pdf

Accesso chiuso

Descrizione: Supplementary material
Tipologia: Altro materiale allegato
Licenza: Copyright Editore
Dimensione 891.08 kB
Formato Adobe PDF
891.08 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11368/3002246
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 11
  • ???jsp.display-item.citation.isi??? 11
social impact